24
(8) Setting for Color-matching Functions
Set and Query Color-matching Functions
Syntax
Command
:ANGLe
<Setting for Color-matching Functions(NRf)>
Query
:ANGLe?
Response
<Setting for Color-matching Functions(NR1)>
<Setting for Color-matching Functions> = 2,10[°]
Description
Command
Sets the color-matching function setting (2° or 10°).
When the command is accepted, the measured value is cleared and the
instrument enters the non-measurement state.
Query
Queries the currently set color-matching function setting.
Example
:ANGL 2
:ANGL?
2
Note
When 10° is set, the instrument operates as follows.
• The photometric quantity is the value that is measured when the color-matching
function setting (2°) is used.
• The correlated color temperature, delta uv, and NTSC ratio cannot be measured. The
instrument enters the non-measurement state.
(9) Dark Measurement Settings
Set and Query Dark Measurement Type
Syntax
Command
:DARK:TYPE
<ALL/FIX >
Query
:DARK:TYPE?
Response
< ALL/FIX >
<ALL> = Executes the dark measurement in all ranges.
<FIX> = Executes the dark measurement in only currently set range (set by RANGe:#).
Description
Command
Sets the dark measurement type.
When the command is accepted, the measured value is cleared and the
instrument enters the non-measurement state.
Query
Queries the currently set dark measurement type.
Example
:DARK:TYPE ALL
:DARK:TYPE?
ALL
Note
When the auto range is enabled, the dark measurement is performed in all ranges
regardless of this setting.