22
(7) Measurement Range
Set and Query Auto Range Function (
# is R, G, or B
)
Syntax
Command
:RANGe:AUTO:#
<1/0/ON/OFF >
Query
:RANGe:AUTO:#?
Response
< 1/0 (NR1)>
Description
Command
Sets the auto range function.
When the command is accepted, the measured value is cleared and the
instrument enters the non-measurement state.
Query
Queries the currently set auto range function.
Example
:RANG:AUTO:R ON
:RANG:AUTO:R?
1
:RANG:AUTO:B OFF
:RANG:AUTO:B?
0
Note
• # is R, G, or B.
• When auto range function is turned off during a range search, the measurement range
may become the impertinent range.When you turn off an automatic range during the
measurement, please set measurement range by all means.
Set and Query Measurement Range (
# is R, G, or B
)
Syntax
Command
:RANGe:#
<Measurement range number(NRf)>
Query
:RANGe:#?
Response
<Measurement range number(NR1)>
<Measurement range number> = 1 to 16
Description
Command
Sets the measurement range.
A larger number increases the sensitivity. (A weak light can be measured.)
When the command is accepted, the measured value is cleared and the
instrument enters the non-measurement state.
Query
Queries the currently set range.
Example
:RANG:R 15
:RANG:R?
15
:RANG:B 16
:RANG:B?
16
Note
•# is R, G, or B.
• A larger number increases the sensitivity. (A weak light can be measured.)
• When the measurement range is changed in the auto range ON state, the auto range
function of the changed color enters the OFF state.