![Hioki SW9001 Instruction Manual Download Page 44](http://html1.mh-extra.com/html/hioki/sw9001/sw9001_instruction-manual_2137434044.webp)
40
Setting the Scan Channel
4.2 Setting the Scan Channel
The channels to be scanned need to be registered as a scan list.
Specify the start and end channels. An individual channel list can also be specified.
Item
Communication command
Setting
procedure
[:ROUTe]:SCAN<(@
Channel list
)>
<(@
Channel list
)>
Channel list format
<CH>,<CH>,...,<CH>
Specify the channels to be measured individually.
<CHm>:<CHn>
Specify from CHm to CHn consecutively.
Setting
example
Scan CH1 of
SLOT 1
to CH22 of
SLOT 3
.
:SCAN 101:322
Scan CH1 and CH2 of
SLOT 1
and CH1, CH2, and CH5 of
SLOT 2
.
:SCAN (@101,102,201,202,205)
4.3 Setting the Trigger Source for Scan Operation
A trigger source for scan operation is set for STEP only.
Trigger source
Scan operation
STEP
Scanning is started by the SCAN input signal or
*TRG
command (the first channel is
closed). Then the next scan channel is scanned every time the SCAN input signal or
*TRG
command is sent.
STEP is set as the default setting. Normally select STEP and execute scanning while
synchronizing with control of the instrument.
The commands to be set are as follows.
Item
Communication command
Setting
procedure
:TRIGger:SOURce <STEP>
<STEP>
STEP:
External trigger. Scanning is performed with the SCAN input signal of EXT. I/O or the
*TRG
command. Scan operation moves along the steps every time the trigger is input.
Setting
example
Set to proceed with the next step every time the external trigger is input.
:TRIG:SOUR STEP
HIOKI SW1001A961-04
Summary of Contents for SW9001
Page 2: ...HIOKI SW1001A961 04...
Page 42: ...38 Channel Delay Function HIOKI SW1001A961 04...
Page 48: ...44 Scan Measurement Example HIOKI SW1001A961 04...
Page 50: ...46 Checking the Device Status HIOKI SW1001A961 04...
Page 52: ...48 Initialization Settings HIOKI SW1001A961 04...
Page 118: ...114 Sample Programs 3 Select FILE Save All HIOKI SW1001A961 04...
Page 160: ...156 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2 HIOKI SW1001A961 04...
Page 161: ...157 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix HIOKI SW1001A961 04...
Page 162: ...158 Outline Drawings HIOKI SW1001A961 04...
Page 164: ...160 HIOKI SW1001A961 04...
Page 167: ...HIOKI SW1001A961 04...
Page 168: ...HIOKI SW1001A961 04...
Page 169: ...HIOKI SW1001A961 04...