94
Message Reference
Device-specific commands
(1) Event Status Register
Query Condition Register of Standard Operation Register Group
Syntax
Query
:STATus:OPERation:CONDition?
Response
<0 to 65535 (NR1)>
Description
Returns the present standard operation register value. Unused bits (indicated in the
following chart by “–”) return the value 0.
bit 15
bit 14
bit 13
bit 12
bit 11
bit 10
bit 9
bit 8
–
–
ERR
–
CLOSE
REMOTE
–
–
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
–
–
WAIT_
TRG
SCAN
–
–
–
–
Example
:STAT:OPER:COND?
1024
The device is in the remote state.
Note
• The present status is returned when this command is received.
•
Unlike
:STATus:OPERation:EVENt?
, :
STATus:OPERation?
, bit information is not
retained.
Query Event Register of Standard Operation Register Group
Syntax
Query
:STATus:OPERation[:EVENt]?
Response
<0 to 65535 (NR1)>
Description
Returns the retained standard operation register value. Unused bits (indicated in the
following chart by “–”) return the value 0.
bit 15
bit 14
bit 13
bit 12
bit 11
bit 10
bit 9
bit 8
–
–
ERR
–
CLOSE
REMOTE
–
–
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
–
–
WAIT_
TRG
SCAN
–
–
–
–
Example
:STAT:OPER:EVEN?
1024
The device is in the remote state.
Note
• When an event occurs, 1 is set for the bit corresponding to each event and retained.
•
The bit is not cleared until
*
CLS
or this query is executed or the power is turned on again.
Set and query Enable Register of Standard Operation Register Group
Syntax
Command
:STATus:OPERation:ENABle <0
-
65535(NR1)>
Query
:STATus:OPERation:ENABle?
Response
<0 to 65535 (NR1)>
Description
Sets the SRER mask pattern to a value from 0 to 255. The value when the device is
turned on is 0. When issued as a query, unused bits are always set to 0.
bit 15
bit 14
bit 13
bit 12
bit 11
bit 10
bit 9
bit 8
–
–
ERR
–
CLOSE
REMOTE
–
–
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
–
–
WAIT_
TRG
SCAN
–
–
–
–
Note
The initial value when the power is turned on is 0.
Summary of Contents for SW1001
Page 2: ......
Page 42: ...38 Precautions for Measurement...
Page 48: ...44 Resetting Scan Operation...
Page 50: ...46 Checking the Device Status...
Page 52: ...48 Initialization Settings...
Page 118: ...114 Sample Programs 3 Select File Save All 4 Check the save location and click Save as...
Page 156: ...152 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2...
Page 157: ...153 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix...
Page 158: ...154 Outline Drawings...
Page 161: ...16 01 EN...
Page 162: ......
Page 163: ......
Page 164: ......