68
RS-232C Interface
Setting RS-232C communications
Set the commands shown below for the transmission speed.
:SYSTem:COMMunicate:RS232C:SPEED <9600/19200/38400>
The speed can be selected from 9600 bps, 19200 bps, or 38400 bps.
The speed setting is immediately changed.
Other RS-232C communication parameters are fixed to 8 for data bit length, 1 for stop bit, none for
parity check, and none for flow control.
When the communication setting mode switch (
DFLT
/
USER
) is set to
DFLT
, the speed setting is
fixed to 9600 bps regardless of the settings specified here.
See: "2.6 Setting the Communication Setting Mode" (p. 22)
Setting the controller (PC, PLC, etc.)
Be sure to set the controller as follows.
Asynchronous (Start-stop) transmission system
Transmission speed
9600 bps/19200 bps/38400 bps
(Set the speed based on the device settings.)
Data bit length
8
Stop bit
1
Parity check
None
Flow control
None
Protocol
Non-procedure
IMPORTANT
Communications may not be established at the high transmission speed (baud rate) depending
on the PC. Lower the transmission speed in that case.
Summary of Contents for SW1001
Page 2: ......
Page 42: ...38 Precautions for Measurement...
Page 48: ...44 Resetting Scan Operation...
Page 50: ...46 Checking the Device Status...
Page 52: ...48 Initialization Settings...
Page 118: ...114 Sample Programs 3 Select File Save All 4 Check the save location and click Save as...
Page 156: ...152 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2...
Page 157: ...153 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix...
Page 158: ...154 Outline Drawings...
Page 161: ...16 01 EN...
Page 162: ......
Page 163: ......
Page 164: ......