LAN Interface
Setting LAN communications
Set the commands shown below for the LAN communications.
:SYSTem:COMMunicate:LAN:IPADdress <Value 1>,<Value 2>,<Value 3>,<Value 4>
Set the IP address for the device.
:SYSTem:COMMunicate:LAN:SMASK <Value 1>,<Value 2>,<Value 3>,<Value 4>
Set the LAN subnet mask.
:SYSTem:COMMunicate:LAN:GATeway <Value 1>,<Value 2>,<Value 3>,<Value 4>
Set the address for the default gateway.
:SYSTem:COMMunicate:LAN:CONTrol <1 - 9999>
Specify the communication command port No.
:SYSTem:COMMunicate:LAN:UPDate
Update and apply the LAN settings.
When the communication setting mode switch (
DFLT
/
USER
) is set to
DFLT
, the default settings are
used regardless of the settings specified here.
Reference: "2.6 Setting the Communication Setting Mode" (p. 22)
Setting items
IP address
An address to identify individual instruments that are connected on the network. Be
sure to set an address that does not overlap with other devices.
Subnet mask
This setting divides the IP address into the address part showing the network and
the address part showing the device. Be sure to set a subnet mask that matches
the subnets of other devices within the same network.
Gateway
IP address
For network connection
When the communication PC and the device are in different networks, specify the
IP address of the device that becomes a gateway.
If the PC is in the same network, set a gateway address that is the same as the
default gateway in the general PC settings.
For one-to-one connection between the device and PC or when no gateway
is used
Set the IP address to
0.0.0.0
.
Communication
command port No.
Specify the TCP/IP port No. used for connecting communication commands.
Summary of Contents for SW1001
Page 2: ......
Page 42: ...38 Precautions for Measurement...
Page 48: ...44 Resetting Scan Operation...
Page 50: ...46 Checking the Device Status...
Page 52: ...48 Initialization Settings...
Page 118: ...114 Sample Programs 3 Select File Save All 4 Check the save location and click Save as...
Page 156: ...152 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2...
Page 157: ...153 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix...
Page 158: ...154 Outline Drawings...
Page 161: ...16 01 EN...
Page 162: ......
Page 163: ......
Page 164: ......