Accuracy Calculation Example
11.3
Accuracy Calculation Example
This section describes how to calculate the combined accuracy of the device and a connected
measuring instrument.
This process consists of calculating the measurement influence for each measuring instrument
(target instrument) used in the combination.
The measurement accuracy when using the device and a module is calculated using the following
formula:
Combination measurement accuracy = Instrument ac combination measurement influence
For more information about a given module’s measurement influences, see the module’s
specifications.
•
SW9001
"Measurement influence" (p. 131)
• SW9002
"Measurement influence" (p. 137)
Below shows an example of accuracy calculation.
Example 1: Use of the device in combination with the BT3562
Used module
SW9001
Instrument
BT3562
Ω
V function, 3 m
Ω
range, 6 V range, sampling SLOW
Measurement value
1 m
Ω
, 4 V
Resistance measurement accuracy = BT3562 resistance measurement ac SW9001
measurement influence (BT3562 3 m
Ω
range)
= (±0.5% rdg. ± 10 dgt.) + (±0.1% f.s.)
= (±0.5% × 1 m
Ω
+ 0.0010 m
Ω
) + {±(0.1% × 3.1000 m
Ω
)}
= (±0.0060 m
Ω
) + (±0.0031 m
Ω
)
= ±0.0091 m
Ω
(total error)
Voltage measurement accuracy = BT3562 voltage measurement ac SW9001
measurement influence (all BT3562 voltage ranges)
= (±0.01% rdg. ± 3 dgt.) + (±5 µV)
= (±0.01% × 4 V ± 0.00003 V) + (±5 µV)
= (±0.00043 V) + (±5 µV)
= ±0.000435 V (total error)
11
Multiplexer Module
Summary of Contents for SW1001
Page 2: ......
Page 42: ...38 Precautions for Measurement...
Page 48: ...44 Resetting Scan Operation...
Page 50: ...46 Checking the Device Status...
Page 52: ...48 Initialization Settings...
Page 118: ...114 Sample Programs 3 Select File Save All 4 Check the save location and click Save as...
Page 156: ...152 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2...
Page 157: ...153 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix...
Page 158: ...154 Outline Drawings...
Page 161: ...16 01 EN...
Page 162: ......
Page 163: ......
Page 164: ......