Gradual Voltage Rise Test
7.3 Gradual Voltage Rise Test
Setting the test conditions
Setting the test method
1
1
1
Press
[BASIC]
>
[BDV MODE]
to set the test
method to
[STEP]
.
(When
[BDV MODE]
is set to
[RATE]
)
A confirmation message is displayed.
2
2
2
Press
[YES]
.
The output settings are initialized.
Setting the initial voltage
The initial voltage is set.
Press the
START
button and wait till the voltage reaches the initial voltage after 0.1 seconds.
Perform the test according to the set voltage holding time, voltage step, and number of steps.
1
1
1
Press
[BASIC]
>
[START VOLT]
on the setting
screen.
2
2
2
Change the value using
[▲]
or
[▼]
and then
press
[ENTER]
.
0.25 kV
,
0.5 kV
,
1 kV
,
2 kV
, any voltage
Setting range:
0.010 kV
*
1
to
7.999 kV
*1. Default setting