
63
Starting and Ending a Test
Stopping a test
When stopping the test forcefully or ending the test when the test time is set to
[CONT]
, press the
STOP
button. A pass/fail judgment is not made at this time.
Front
STOP
button
Ending a test
The test ends in the following cases.
(1) When a current over the upper-limit judgment reference value or below the lower-limit
judgment reference value is detected
(2) When the test time is not set to
[CONT]
and the test time has elapsed
(3) When the
STOP
button is pressed
When the test ends,
[TEST]
in the display area turns off.
When the output terminals are discharged and the residual voltage in the output terminals becomes
zero, the
DANGER
lamp turns off.
If the test ends in (1) and (2) above, the judgment result is displayed in the display area.
IMPORTANT
If the time setting (such as the test time
[TEST TIME]
, voltage rise time
[RISE TIME]
, voltage fall
time
[FALL TIME]
, interval time
[INTERVAL]
, delay time
[DELAY]
) is set to the shortest time,
and the test is repeated at the fastest time, the display of the screen may be delayed from the
actual measurement. If the screen display is delayed, set a longer time.
Withstand V
oltage
Test/Insulation Resistance
Test Mode
HIOKI ST5680A961-00
Summary of Contents for ST5680
Page 2: ...HIOKI ST5680A961 00 ...
Page 84: ...84 Individually Executing Specific Steps HIOKI ST5680A961 00 ...
Page 118: ...118 Key Lock HIOKI ST5680A961 00 ...
Page 154: ...154 Calibration Expiration Check Function HIOKI ST5680A961 00 ...
Page 182: ...182 Internal Circuit Configuration HIOKI ST5680A961 00 ...
Page 220: ...220 Option Specifications HIOKI ST5680A961 00 ...
Page 238: ...238 External View 16 4 External View Unit mm HIOKI ST5680A961 00 ...
Page 250: ...250 HIOKI ST5680A961 00 ...
Page 253: ...www hioki com HIOKI ST5680A961 00 ...