50
Setting the Voltage Rise Time
4.10 Setting the Voltage Rise Time
The time period from the start of a test until the test voltage is reached or from the start voltage until
the test voltage is reached can be set.
DCW setting
1
1
1
Press
[BASIC]
>
[RISE TIME]
on the setting
screen.
2
2
2
Change the value using
[▲]
or
[▼]
and then
press
[ENTER]
.
0.1 s
*
1
to
300 s
*1. Default setting
The lower limit is set by
[TEST TIME]
(test time) and
[DELAY]
(judgment wait time).
[DELAY]
<
[RISE TIME]
+
[TEST TIME]
(Only when
[DELAY]
is not set to off)
IR setting
1
1
1
Press
[BASIC]
>
[RISE TIME]
on the setting
screen.
2
2
2
Change the value using
[▲]
or
[▼]
and then
press
[ENTER]
.
0.1 s
*
2
to
300 s
*2. Default setting
The lower limit is set by
[TEST TIME]
(test time) and
[DELAY]
(judgment wait time).
[DELAY]
<
[RISE TIME]
+
[TEST TIME]
(Only when
[DELAY]
is not set to off)