Timer Calibration
16.6 Timer Calibration
Calibration conditions
Ambient temperature
and humidity
23
°
C
±
5
°
C, 80% RH or less
Warm-up time
30 minutes or more
Power supply
100 V to 240 V 50 Hz or 60 Hz
Caliblation equipment
Frequency counter
Keysight 53220A or equivalent
Connection
Use the TEST signal (19 pin) in the external I/O terminal.
External I/O terminal
19 TEST
8 ISO_5 V
Input
Keysight
53220A
ST5680
ST5680 settings
Test mode
DCW or IR
Test time
5.0 s
Output voltage
1.000 kV
Load not connected (open)
TEST signal outputting timing
TEST (Output the singal during the test time)
See “Setting the test signal output timing” (p. 162)
Press the
START
button to start a test and measure the pulse width of the TEST signal.
HIOKI ST5680A961-00