196
Control Using Communications Commands
13.8 Control Using Communications Commands
For the communications command and query notations (communication message reference), see
the Communications Command Instruction Manual on the provided CD.
Remote state
If starting communications after connecting an interface to the instrument, the instrument enters the
remote state (remote operations mode) and
[REMOTE]
is displayed on the measurement screen.
In the remote state, touch screen operations other than
[LOCAL]
become disabled.
Local state
The following operations reset the remote state and allow the instrument to enter the local state that
enables touch screen operation.
• Press
[LOCAL]
.
• Turn on the instrument again.
• Send the
:SYSTem:LOCal
command to the instrument via a communication interface.
• Send the
GTL
command to the instrument from GP-IB.