Switching the Measurement Speed (FAST/SLOW)
3.3 Switching the Measurement Speed (FAST/
SLOW)
There are two measurement speeds,
[FAST]
and
[SLOW]
. The resistance is
measured and displayed in 30 ms when the measurement speed is set to
[FAST]
and
in 500 ms when the speed is set to
[SLOW]
.
Setting the measurement speed to
[SLOW]
is effective if a measured value fluctuates
and it is difficult to read the value.
1
Open the Settings screen.
Select
[SET]
.
2
Set the test speed.
Selection
[SLOW]
[FAST]
Return to the
Measurement
screen.
IMPORTANT
•
The sampling time is 100 ms even when the measurement speed is set to
[FAST]
if the contact check function is ON.
(The sampling time is 500 ms if the measurement speed is set to
[SLOW]
regardless of the function check setting.)
•
If the test duration is set to less than 500 ms when the measurement speed is set
to
[SLOW]
, the measured value is not displayed. Set a test duration that is longer
than the sampling time.