36
Setting the Range
3.2 Setting the Range
There are two range settings, auto and manual. For the manual range setting, the
range can be chosen from 2 M
Ω
, 20 M
Ω
, 200 M
Ω
, 2000 M
Ω
, and 4000 M
Ω
. Note,
however, that the ranges that can be set varies depending on the test voltage. (p. 153)
The auto range may require some time before a measured value is displayed as the
range has to shift after a test starts. To eliminate this time, set the range manually
according to the object to be measured.
Setting the manual range
1
Open the Settings screen.
Select
[SET]
.
2
Set
[RANGE]
to
[HOLD]
.
Selection
[HOLD]
Return to the
Measurement
screen.
3
Select the range.
Item selection
Select the range.
Return to the
Measurement
screen.
When analog output is used, 0 to 4 V is output from the analog output terminal according to