140
Command Reference
:SHORtcheck:TIME?
Queries the short circuit check time.
The short circuit check time is returned in NR2 numerical data (<data>). 0.000 is returned when
the short circuit check time is set to auto (AUTO).
Syntax
:SHORtcheck:TIME?
Response
:SHORTCHECK:TIME <data>
or simply
<data>
(NR2 numerical data)
<data>
Short circuit check time (unit: second) 0.000 and 0.010 to 1.000 (resolution
0.001)
Example
To query the short circuit check time (when the short circuit check
execution time is 0.017 seconds)
PC>
:SHORtcheck:TIME?
ST5520>
:SHORTCHECK:TIME 0.017
(when
the
:HEADer
command is ON)
ST5520>
0.017
(when the
:HEADer
command is OFF)
:SHORtcheck:TIME:MONItor?
The short circuit check execution time when the short circuit check time is set to 0.0 (AUTO) is
returned in NR2 numerical data (<data>). 0.000 is returned when the short circuit check function
is set to OFF or MANUAL.
No command header is included in the response.
Syntax
:SHORtcheck:TIME:MONItor?
Response
<data>
(NR2 numerical data)
<data>
Short circuit check execution time (unit: second) 0.000 to 1.000 (resolution
0.001)
Example
To query the short circuit check execution time (when the short circuit
check execution time is 0.017 seconds)
PC>
:SHORtcheck:TIME:MONItor?
ST5520>
0.017
:SHORtcheck:RESult?
Queries the short circuit check result.
The result is returned in text data (<data>).
No command header is included in the response.
Syntax
:SHORtcheck:RESult?
Response
<data>
<data>
NOCHK/PASS/FAIL (text data)
NOCHK: Not checked yet
PASS: No error
FAIL: Short circuit error
Example
To query the short circuit check result (when the result is a short circuit
error)
PC>
:SHORtcheck:RESult?
ST5520>
FAIL
Summary of Contents for ST5520
Page 2: ......
Page 8: ...vi Contents ...
Page 22: ...14 Usage Notes ...
Page 56: ...48 Announcing the Judgment Results Using a Beep Sound ...
Page 82: ...74 Initializing the System Reset ...
Page 88: ...80 Deleting the Panel Data ...
Page 158: ...150 Command Reference ...
Page 193: ...Appx 13 Rack Mounting Screw location dimensional drawing 10 9 8 7 6 5 4 3 2 1 Appx Ind ...
Page 194: ...Appx 14 Dimensional Diagram Appx 8 Dimensional Diagram 166 2 215 1 14 1 80 1 ...
Page 197: ...13 09 ...
Page 198: ......
Page 199: ......
Page 200: ......
Page 201: ......