Command Reference
:SHORtcheck
Sets the short circuit check function ON/OFF.
Syntax
:SHORtcheck <data>
<data>
ON/OFF (text data)
ON: The short circuit check function is enabled.
OFF: The short circuit check function is disabled.
Example
To set the short circuit check function to ON
PC>
:SHORtcheck ON
:SHORtcheck?
Queries the short circuit check function ON/OFF setting. The setting is returned in text data (<data>).
Syntax
:SHORtcheck?
Response
:SHORTCHECK <data>
or simply
<data>
<data>
ON/OFF (text data)
ON: The short circuit check function is enabled.
OFF: The short circuit check function is disabled.
Example
To query the short circuit check function (when the short circuit check
function is enabled)
PC>
:SHORtcheck?
ST5520>
:SHORTCHECK ON
(when the
:HEADer
command is ON)
ST5520>
ON
(when the
:HEADer
command is OFF)
:SHORtcheck:TIME
Sets the short circuit check time.
Set 0.0 when the short circuit check time is set to auto (AUTO).
Syntax
:SHORtcheck:TIME <data>
<data>
Short circuit check time (unit: second) 0.0 and 0.010 to 1.000 (resolution
0.001) (NR2 numerical data)
Example
To set the short circuit check time to 0.5 seconds
PC>
:SHORtcheck:TIME 0.5
10
9
8
7
6
5
4
3
2
1
Appx.
Ind.
Summary of Contents for ST5520
Page 2: ......
Page 8: ...vi Contents ...
Page 22: ...14 Usage Notes ...
Page 56: ...48 Announcing the Judgment Results Using a Beep Sound ...
Page 82: ...74 Initializing the System Reset ...
Page 88: ...80 Deleting the Panel Data ...
Page 158: ...150 Command Reference ...
Page 193: ...Appx 13 Rack Mounting Screw location dimensional drawing 10 9 8 7 6 5 4 3 2 1 Appx Ind ...
Page 194: ...Appx 14 Dimensional Diagram Appx 8 Dimensional Diagram 166 2 215 1 14 1 80 1 ...
Page 197: ...13 09 ...
Page 198: ......
Page 199: ......
Page 200: ......
Page 201: ......