
73
Discharge Judgments (When Incorporating Model ST9000)
Discharge judgment methods
[DISCHARGE]
(Measurement screen)
[MODE]
>
[SETTING]
>
[JUDGE]
>
[DISCHARGE]
1
1
Tap
[ENABLE]
, and turn
[ON]
the discharge judgment methods.
OFF
Disables the Discharge Detection Upgrade.
FIXED
Compares the amount of the discharge component of the master waveform and the amount of
the discharge component of the test waveform.
AUTO
Calculates variations in the high-frequency component that appears in test waveforms as the
standard deviation (
σ
). Discharges will be detected based on whether the deviation exceeds a
fixed amount due to a sudden increase in the high-frequency component due to discharge.
IMPORTANT
If the discharge judgment method is AUTO, suitable filtering is automatically selected according the the
response waveform of the conforming workpiece by implementing voltage calibration. The discharge
judgment method AUTO has a higher discharge judgment performance than FIXED, so normally make sure
to select the AUTO setting. If using a special workpiece that emits a greatly different response waveform
than normal, it may be difficult to render judgments when the discharge judgment method is AUTO. In such
cases, set the discharge judgment method to FIXED.
5
Setting Judgment Conditions
Summary of Contents for ST4030
Page 2: ......
Page 14: ...6 Usage Notes ...
Page 32: ...24 Inspecting before Measurement and Verifying Operations ...
Page 42: ...34 Table Copying ...
Page 108: ...100 Setting Dielectric Break down Judgment Conditions ...
Page 132: ...124 Terminal Open Error Setting ...
Page 198: ...190 Editing Files and Folders ...
Page 214: ...206 Interface Specifications ...
Page 248: ...240 Precautions when Processing Test Leads ...
Page 252: ...244 Index ...
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......