69
Waveform Judgments
Secondary differential detection judgments
[LAPLACIAN]
The amount of the high-frequency component expressed by the test waveform in the user-
designated area is detected, and the size of the amount used to judge the discharge. The response
waveform is secondarily differentiated and the surface area of the differentiated waveform
calculated as the amount of the high-frequency component.
Laplacian operations
Laplacian
=
Number of data
−
1
Σ
n
=1
|−
d
n
−1
+2
d
n
−
d
n
+1
|
Enabling secondary differential detection judgment functions
(Measurement screen)
[MODE]
>
[SETTING]
>
[JUDGE]
>
[LAPLACIAN]
1
1
Tap
[ENABLE]
, and turn
[ON]
the secondary differential detection judgments.