64
Waveform Judgments
Differential surface area comparison judgments
[DIFF AREA]
Calculates the surface area surrounded by the master waveform and test waveform in the user-
designated area to judge whether the test workpiece within that area conforms.
The surface area surrounded by the master and test waveforms in the judgment area (surface area
in a solid orange color) is calculated as shown in the “Differential surface area calculation example”
below.
The judgment standards are set using percentages, and the test workpiece is judged to conform if
the percentage of the test waveform surface area over the master waveform surface area is in that
range.
Differential surface area calculation example
Judgment area
Differential surface area comparison value operation formula
Diff
(%) =
(Test waveform surface area value −
Master waveform surface area value)
Master waveform surface area value
*
100
Enabling the differential surface area comparison judgment function
(Measurement screen)
[MODE]
>
[SETTING]
>
[JUDGE]
>
[DIFF AREA]
1
1
Tap
[ENABLE]
, and turn
[ON]
the differential surface area comparison judgment function.
Summary of Contents for ST4030
Page 2: ......
Page 14: ...6 Usage Notes ...
Page 32: ...24 Inspecting before Measurement and Verifying Operations ...
Page 42: ...34 Table Copying ...
Page 108: ...100 Setting Dielectric Break down Judgment Conditions ...
Page 132: ...124 Terminal Open Error Setting ...
Page 198: ...190 Editing Files and Folders ...
Page 214: ...206 Interface Specifications ...
Page 248: ...240 Precautions when Processing Test Leads ...
Page 252: ...244 Index ...
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......