Waveform Differences (Typical Values) Between Instruments
15.8 Waveform Differences (Typical Values) Between
Instruments
Even a slight difference in the interior of the impulse winding tester produces a difference in the
oscillation waveform even if using the same workpiece and test leads.
Instrument differences for the voltage peak value and cycle (zero-cross value) obtained with 14
Model ST4030 instruments when using the same workpiece and test leads are described as
reference values.
Test conditions
Workpiece
10 µH coil
1 mH coil
Sampling frequency
200 MHz
200 MHz
Sampling data count
1001 points
5001 points
Applied voltage
1000 V
Voltage calibration
Implemented
Test lead
L2250
Voltage [V]
First peak voltage
Second peak voltage
Third peak voltage
Time [µs]
First wave
half-period
Second
wave period
Describes the instrument differences determined using the waveforms obtained using each Model
ST4030.
Formula: (Max. value − min. value) / average value [%]
Workpiece
10 µH coil
1 mH coil
First peak voltage
0.90%
0.06%
Second peak voltage
1.80%
0.45%
Third peak voltage
4.46%
0.50%
First wave half-period
0.87%
0.29%
Second wave period
1.09%
0.37%
Summary of Contents for ST4030
Page 2: ......
Page 14: ...6 Usage Notes ...
Page 32: ...24 Inspecting before Measurement and Verifying Operations ...
Page 42: ...34 Table Copying ...
Page 108: ...100 Setting Dielectric Break down Judgment Conditions ...
Page 132: ...124 Terminal Open Error Setting ...
Page 198: ...190 Editing Files and Folders ...
Page 214: ...206 Interface Specifications ...
Page 248: ...240 Precautions when Processing Test Leads ...
Page 252: ...244 Index ...
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......