231
Repeat Accuracy (Reference Value)
Oscillation waveform reproducibility
Describes the typical values of waveform variation when implementing impulse tests on the coil.
The oscillation waveform when implementing impulse tests on a low-loss 1 mH coil using 1000 V is
as described below.
Voltage [V]
First peak voltage
Second peak voltage
Third peak voltage
Time [µs]
The
repeat accuracy of the peak voltages when implementing 100 consecutive tests is as described
below.
Nearly all 0.3% max.
Repeat accuracy [%]
First peak
Second peak
Third peakFourth peak Fifth peakSixth peak
Seventh peak
Note: The values shown here are typical values, and repeat accuracy cannot be assured.
Waveform variation depends greatly on the inductance, stray capacitance, linear resistance, and
core magnetic characteristics of the workpiece being tested.
For example, the peak voltage will drift a little more each time due to the increasing residual
magnetization of the magnetic core.
For good impulse test reproducibility, implement degaussing pulses a sufficient number of times so
that there is no more waveform drift.
15
Appendix
Summary of Contents for ST4030
Page 2: ......
Page 14: ...6 Usage Notes ...
Page 32: ...24 Inspecting before Measurement and Verifying Operations ...
Page 42: ...34 Table Copying ...
Page 108: ...100 Setting Dielectric Break down Judgment Conditions ...
Page 132: ...124 Terminal Open Error Setting ...
Page 198: ...190 Editing Files and Folders ...
Page 214: ...206 Interface Specifications ...
Page 248: ...240 Precautions when Processing Test Leads ...
Page 252: ...244 Index ...
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......