126
Self-test Function
9.2
Self-test Function
You can self-test for errors in the instrument.
Touch panel test
You can check that the touch panel is operating normally.
(Measurement screen)
[SYSTEM]
>
[TEST]
1
2
3
1
Tap
[EXEC]
.
2
Tap each black square in turn.
(Black)
Testing is not implemented.
(Gray)
You can tap squares that are normally black.
(Red)
Tapping other than black squares. If you cannot tap a black square even after trying
several times, implement touch panel compensation.
Reference: “Touch panel correction” (p. 127)
IMPORTANT
If the black square still cannot be tapped even after panel correction, there is a risk of malfunction. Contact
your authorized Hioki distributor or reseller.
3
When the test is finished, tap
[EXIT]
.