3.4 Configuring the Communications Protocol
21
Setting the address
Communication conditions
Remote switching requests
Send the “
RMT
” command from the RS-232C interface.
3.4
Configuring the Communications Protocol
Configuring GP-IB Interface Communications
1
Press and hold the scroll keys (
PAGE UP
▲
/
PAGE DOWN
▼
) on the front of the device
for about 7 seconds. (The address can be set from the P1 or P2 screen.)
■
Screen P1: Status display
3
When finished making the setting, turn off the device.
4
Turn on the device.
The device will revert to the initial screen, and the GP-IB address will be set to the
selected address.
G P I B A D D R : 1
V A ( + ) I R : 1 0 0 0 . 0
V B ( + ) I R :
2 5 0 . 0
T E R M I N A L :
O U T 1 : 1 1 1 0 0 0 0 0
O U T 2 : 1 1 1 0 0 0 0 0
O U T 3 : 1 1 1 0 0 0 0 0
O U T 4 : 1 1 1 0 0 0 0 0
V M A : 1 0 0 0 . 0
O K
V M B :
2 5 0 . 0
O K
2
Using the scroll keys (
PAGE UP
▲
/
PAGE DOWN
▼
), set the desired address.
(Valid setting range: 1 to 30)
Configuring RS-232C Interface Communications
Baud rate
38400 bps
Parity
none
Stop bit
1 bit
Data
8 bits
Flow control
none
RS-232C interface communications will not be available until the “
RMT
” com-
mand is sent.