Appendix 15 Using the Instrument with a Withstanding Voltage Tester
A33
A
ppend
ix
The instrument can also be used in conjunction with a withstanding voltage tester to test
windings. When used with a withstanding voltage tester, the charge stored in the winding
may flow into the instrument at the moment it is connected, damaging it. When using the
instrument in this manner, take the following into account during the production line design
process:
(1) Ensure the contact withstanding voltage of the relays used for switching has a sufficient
safety margin relative to the withstanding test voltage (at a minimum, it should be twice
the peak voltage).
Example high-voltage relays
Okita Works
LRL-101-50PC (5 kV DC between contacts)
LRL-101-100PC (10 kV DC between contacts)
Sanyu Switch
USM-11524 (5 kV DC between contacts)
USM-13624SB (10 kV DC between contacts)
(2) During withstanding voltage testing, ground all of the instrument’s terminals.
(3) Perform resistance measurement first and the withstanding voltage test last.
If you must perform the withstanding voltage test before resistance measurement, ground
both of the measurement target’s terminals after the withstanding voltage test to discharge
any charge accumulated during the test. Then perform resistance measurement.
Appendix 15 Using the Instrument with a
Withstanding Voltage Tester
Ground measurement terminals
when not performing resistance measurement.
Instrument
Withstanding
voltage tester
Discharge residual
charge.
Measurement target
Use only high-voltage relays.
Using the instrument with a withstanding voltage tester
Summary of Contents for RM3545
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 18...
Page 42: ...1 5 Checking the Measurement Target 34...
Page 54: ...2 6 Pre Operation Inspection 46...
Page 70: ...3 5 Checking Measured Values 62...
Page 226: ...10 6 Supplied Connector Assembly 218...
Page 290: ...Chapter 13 Specifications 282...
Page 312: ...14 4 Disposing of the Instrument 304...
Page 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Page 362: ...Index Index 4...
Page 363: ......
Page 364: ......