14.4 Disposing of the Instrument
302
The instrument uses a lithium battery for back-up power to the clock.
When disposing of this instrument, remove the lithium battery and dispose of battery and
instrument in accordance with local regulations.
14.4 Disposing of the Instrument
Removing the Lithium Battery
To avoid electric shock, turn off the main power button and disconnect
the power cord and measurement leads before removing the lithium
battery.
Required tools:
•
One Phillips screwdriver (No.1)
•
One wire cutter (to remove the lithium battery)
1
Remove the six screws from the sides
and one screw from the rear.
2
Remove the cover.
Insert the tweezers between the battery
and battery holder as shown in the dia-
gram below and lift up the battery.
3
Verify that the power is off, and remove
the connection cables and power cord.
4
(Overhead View)
Take care not to short the + and -. Doing so
may cause sparks.
CALIFORNIA, USA ONLY
This product contains a CR Coin Lithium Battery which contains
Perchlorate Material - special handling may apply.
See www.dtsc.ca.gov/hazardouswaste/perchlorate
RM3545, RM3545-01
Battery Holder
Lithium
Battery
Summary of Contents for RM3545
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 18...
Page 42: ...1 5 Checking the Measurement Target 34...
Page 54: ...2 6 Pre Operation Inspection 46...
Page 70: ...3 5 Checking Measured Values 62...
Page 226: ...10 6 Supplied Connector Assembly 218...
Page 290: ...Chapter 13 Specifications 282...
Page 312: ...14 4 Disposing of the Instrument 304...
Page 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Page 362: ...Index Index 4...
Page 363: ......
Page 364: ......