10.4 External I/O Settings
207
10
Switching the trigger source
Continuous measurement (
:INITIATE:CONTINUOUS ON
) is the normal trigger state when
using key operation from the front panel. Selecting the internal (INT) trigger source acti-
vates continuous triggering ("free-run"). When external (EXT) triggering is selected, each
external trigger event initiates one measurement. Continuous measurement can be dis-
abled by sending the
:INITIATE:CONTINUOUS OFF
command via RS-232C, USB or GP-
IB. When continuous measurement is disabled, trigger acceptance is controlled only by the
controller (computer or PLC).
See:
For trigger command: See the included application disc.
1
Open the Settings Screen.
2
Open the EXT I/O Setting Screen.
3
Select the trigger source.
4
Return to the Measurement screen.
The Settings screen
appears.
Switch the function menu
to P.2/3.
1
2
Move the cursor to the [I/O]
tab with the left and right cur-
sor keys.
2
(INT) Internal trigger
(default)
(EXT) External trigger
Selection
1
Return to the
Measurement screen.
Summary of Contents for RM3545
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 18...
Page 42: ...1 5 Checking the Measurement Target 34...
Page 54: ...2 6 Pre Operation Inspection 46...
Page 70: ...3 5 Checking Measured Values 62...
Page 226: ...10 6 Supplied Connector Assembly 218...
Page 290: ...Chapter 13 Specifications 282...
Page 312: ...14 4 Disposing of the Instrument 304...
Page 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Page 362: ...Index Index 4...
Page 363: ......
Page 364: ......