10.2 Timing Chart
200
This section describes the process from measurement start to acquisition of judgment
results or measured values when using an external trigger.
The instrument outputs the EOM signal immediately once the judgment result (HI, IN, LO,
ERR, T_PASS, T_FAIL, T_ERR) has been finalized. If the controller’s input circuit response
is slow, it may be necessary to insert wait processing after the EOM signal’s changing to
ON is detected until the judgment result is acquired.
Acquisition Process When Using an External Trigger
Instrument
Controller
Measurement
start
Judgment re-
sults OFF
Measurement
start
Measurement
Measurement
complete
Reception of
measurement
complete
Acquisition of
judgment results
Wait
Wait for EOM
(when detect-
ing the level,
after a 0.5 ms
wait)
HI, IN, LO, ERR,
T_PASS, T_FAIL,
T_ERR