8.3 Multiplexer Settings
152
The Multiplexer Unit can measure the resistance between user-specified pin pairs by
changing the channel pin allocation. Pin allocation can be set for up to 42 channels.
If you wish to initialize the multiplexer channel settings
See:
"7.7 Initializing (Reset)"(p.134)
Channel default settings
4-wire
2-wire
See:
"8.7 Example Connections and Settings"(p.167)
Customizing Channel Pin Allocation
CH
UNIT
TERM A
TERM B
1
ON
1
TERM1
TERM2
2
OFF
1
TERM2
TERM3
3
OFF
1
TERM3
TERM4
4
OFF
1
TERM4
TERM5
5
OFF
1
TERM5
TERM6
6
OFF
1
TERM6
TERM7
7
OFF
1
TERM7
TERM8
8
OFF
1
TERM8
TERM9
9
OFF
1
TERM9
TERM10
10
OFF
1
TERM10
TERM11
11
OFF
2
TERM1
TERM2
12
OFF
2
TERM2
TERM3
:
:
:
:
:
20
OFF
2
TERM10
TERM11
21
OFF
1
TERM1
TERM2
22
OFF
1
TERM1
TERM2
:
:
:
:
:
42
OFF
1
TERM1
TERM2
CH
UNIT
TERM A
TERM B
1
ON
1
TERM1
TERM2
2
OFF
1
TERM2
TERM3
3
OFF
1
TERM3
TERM4
4
OFF
1
TERM4
TERM5
5
OFF
1
TERM5
TERM6
:
:
:
:
:
21
OFF
1
TERM21
TERM22
22
OFF
2
TERM1
TERM2
23
OFF
2
TERM2
TERM3
:
:
:
:
:
42
OFF
2
TERM21
TERM22
Summary of Contents for RM3545
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 18...
Page 42: ...1 5 Checking the Measurement Target 34...
Page 54: ...2 6 Pre Operation Inspection 46...
Page 70: ...3 5 Checking Measured Values 62...
Page 226: ...10 6 Supplied Connector Assembly 218...
Page 290: ...Chapter 13 Specifications 282...
Page 312: ...14 4 Disposing of the Instrument 304...
Page 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Page 362: ...Index Index 4...
Page 363: ......
Page 364: ......