139
8
By using the RM3545-02 in combination with the Z3003 Multiplexer Unit, it is possible to
conduct measurements by switching among up to 20 locations (4-wire) or up to 42 locations
(2-wire).
When installing the multiplexer unit, be sure to read "2.4 Installing the Multiplexer
Unit"(p.42).
•
The Z3003 Multiplexer Unit’s contacts use mechanical relays. Since mechanical relays
have a finite service life, programs should be created so as to minimize the switching of
contacts.
Contact service life reference value
4-wire: 50 million cycles, 2-wire: 5 million cycles
•
The self-test function performs short and open tests. Short test measures each pin’s
round-trip wiring resistance in the 2-terminal resistance measurement state and gener-
ates a PASS result if the value is 1
Ω
or less. When using a measurement current of 1 A,
it may not be possible to conduct measurement due to an inability to achieve the 1 A
measurement current, even if the self-test yields a PASS result. If you encounter a cur-
rent fault (
-------
or
OvrRng
display), reduce the wiring resistance and the contact resis-
tance between the measurement target and measurement leads. (p.57)
Multiplexer
Chapter 8
Summary of Contents for RM3545
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 18...
Page 42: ...1 5 Checking the Measurement Target 34...
Page 54: ...2 6 Pre Operation Inspection 46...
Page 70: ...3 5 Checking Measured Values 62...
Page 226: ...10 6 Supplied Connector Assembly 218...
Page 290: ...Chapter 13 Specifications 282...
Page 312: ...14 4 Disposing of the Instrument 304...
Page 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Page 362: ...Index Index 4...
Page 363: ......