7.3 Power Line Frequency Manual Setting
129
7
With the default setting (AUTO), the instrument attempts to automatically detect the line fre-
quency, but manual setting is also available.
•
Unless the line frequency is set correctly, measured values may be unstable.
An error message appears if line noise is high enough to prevent correct frequency
detection (ERR:097 (p. 296)). In that case, set the instrument’s line frequency manually.
•
When the AUTO setting is selected, the line frequency is automatically set to 50 or 60 Hz
when the instrument is turned on or reset.
However, automatic detection is not available when the line frequency changes after turn-
ing power on or resetting.
If the actual line frequency deviates from 50 or 60 Hz, select the closest frequency.
Examples:
If the actual line frequency is 50.8 Hz, select the 50 Hz setting.
If the actual line frequency is 59.3 Hz, select the 60 Hz setting.
7.3
Power Line Frequency Manual Setting
1
Open the Settings Screen.
2
Open the System Setting Screen.
The Settings screen
appears.
Switch the function menu
to P.2/3.
1
2
Move the cursor to the [SYS]
tab with the left and right cur-
sor keys.
Summary of Contents for RM3545
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 18...
Page 42: ...1 5 Checking the Measurement Target 34...
Page 54: ...2 6 Pre Operation Inspection 46...
Page 70: ...3 5 Checking Measured Values 62...
Page 226: ...10 6 Supplied Connector Assembly 218...
Page 290: ...Chapter 13 Specifications 282...
Page 312: ...14 4 Disposing of the Instrument 304...
Page 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Page 362: ...Index Index 4...
Page 363: ......
Page 364: ......