119
6
Current measurement conditions can be saved and loaded using the panel load function
from the key operations, communications commands, or EXT-I/O.
The instrument can save up to 30 sets (panel number: 1 to 30) of measurement conditions
when not using the multiplexer or up to 8 sets (panel number: 31 to 38) when using the mul-
tiplexer. This data is retained even if the instrument is turned off.
Settings that can be saved with the Panel Save function
Saving and Loading Panels
(Saving and Loading
Measurement Conditions)
Chapter 6
•
Panel name
•
Save time and date
•
Resistance range
•
100 M
Ω
high-precision mode
•
Low-Power resistance measurement (LP)
•
Switching measurement currents
•
Measurement speed
•
Zero-adjustment (Loading of these values can be disabled.) (p.122)
•
Averaging
•
Delay
•
Temperature correction (TC)
•
Offset voltage compensation (OVC)
•
Scaling
•
Self-calibration settings
•
Contact Improver
•
Contact check
•
Comparator
•
BIN settings
•
Judgment beeper
•
Auto hold
•
Temperature conversion (
Δ
T)
•
Statistical calculations settings
•
Multiplexer settings (including channels)
Summary of Contents for RM3545
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 18...
Page 42: ...1 5 Checking the Measurement Target 34...
Page 54: ...2 6 Pre Operation Inspection 46...
Page 70: ...3 5 Checking Measured Values 62...
Page 226: ...10 6 Supplied Connector Assembly 218...
Page 290: ...Chapter 13 Specifications 282...
Page 312: ...14 4 Disposing of the Instrument 304...
Page 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Page 362: ...Index Index 4...
Page 363: ......
Page 364: ......