5.1 Judging Measured Values (Comparator Function)
99
5
Before Using the Comparator Function
•
The comparator judgment indicator will function as follows for over-range events
(“
OvrRng
” display) and measurement faults (“
CONTACT TERM
” display or “
- - - - - - -
” dis-
play):
See:
"Confirming Measurement Faults"(p.55)
•
If power is turned off during comparator setting, changes to settings are lost as they
revert to their previous values. To accept the settings, press the
.
Measured value display
Comparator Judgment Indicator
+OvrRng
Hi
-OvrRng
Lo
CONTACT TERM
or
- - - - - - -
Off (no judgment)