
Appendix 3 Unstable Measurement Values
A
6
(6)
DUT Becomes Warm
The maximum applied power to a measurement object by this instrument is determined as follows. The resis-
tance of samples with small thermal capacity can change due to heating. In such cases, enable low-power
measurement.
(7) Ingress of External Noise
Measurement should be conducted as far as possible from
devices emitting electric or magnetic fields such as power
cords, fluorescent lights, solenoid valves and PC displays.
If external noise ingress is a problem, prepare the measure-
ment leads as shown in Fig. 9.
(8) Measuring Transformers and Motors
If noise enters an unconnected terminal of a transformer or if motor rotor moves, measurements may be
unstable due to induced voltage on the measured winding. Pay attention to the treatment of unconnected ter-
minals on transformers or to motor vibration.
(9) Measuring Large Transformers or Motors
When measuring high-inductance (high-Q) DUTs such as large transformers or motors, measured values may
be unstable.
The RM3542 depends on constant current flow through the DUT, but producing constant current becomes
impossible as inductance approaches infinity. To obtain stability in a constant-current source with a large
inductance, response time is sacrificed. If you find that resistance values are scattered when measuring large
transformers or motors, please consider the above or contact your local HIOKI distributor for further assis-
tance.
Range
Measurement
Current
Maximum Applied Power
= (Measured Resistance)
x
(Measurement Current)
2
100 m
100 mA
1.2 mW
1000 m
100 mA
12 mW
10
10 mA
1.2 mW
100
10 mA
12 mW
1000
1 mA
1.2 mW
10 k
1 mA
12 mW
100 k
100
A
1.2 mW
1000 k
10
A
120
W
10 M
1
A
12
W
100 M
100 nA
1.2
W
LP1000 m
10 mA
120
W
LP10
1 mA
12
W
LP100
1 mA
120
W
LP1000
100
A
12
W
Figure 9. Wiring to Minimize Noise Ingress
Twisted
Metal
Cover
Summary of Contents for RM3542
Page 2: ......
Page 6: ...Contents iv ...
Page 16: ...Operating Precautions 10 ...
Page 26: ...1 3 Screen Organization 20 ...
Page 32: ...2 3 Turning the Power On and Off 26 ...
Page 46: ...3 8 Confirming Faulty Measurements 40 ...
Page 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Page 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Page 90: ...7 3 Printing 84 ...
Page 172: ...9 10 Device Compliance Statement 166 ...
Page 190: ...11 4 Disposing of the Instrument 184 ...
Page 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Page 214: ...Index Index 4 ...
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......