Chapter 10 Specifications
177
10
(2) Measurement Time (default settings)
Environment and Safety Specifications
Low Power OFF
Low Power ON
Range
FAST
MED
SLOW
100m
3.8 ms
13 ms
43 ms
36 ms
1000m
2.0 ms
6.4 ms
41 ms
35 ms
10
1.6 ms
6.0 ms
41 ms
34 ms
100
0.9 ms
3.6 ms
21 ms
17 ms
1000
0.9 ms
3.6 ms
21 ms
17 ms
10k
1.0 ms
3.6 ms
21 ms
17 ms
100k
1.3 ms
3.8 ms
21 ms
18 ms
1000k
2.5 ms
6.0 ms
21 ms
18 ms
10M
5.3 ms
23 ms
20 ms
23 ms
20 ms
100M
26 ms
22 ms
46 ms
39 ms
86 ms
72 ms
Upper value: 50Hz power line frequency
Lower value: 60Hz power line frequency
Tolerance ±10%
±
0.2 ms
(Without retry)
Range
FAST
MED
SLOW
1000m
2.5 ms
12 ms
42 ms
35 ms
10
2.5 ms
12 ms
42 ms
35 ms
100
1.7 ms
6.1 ms
41 ms
34 ms
1000
7.2 ms
12 ms
47 ms
40 ms
Upper value: 50Hz power line frequency
Lower value: 60Hz power line frequency
Tolerance ±10%
±
0.2 ms
(Without retry)
Operating environment
Indoors, Pollution degree 2, up to 2000 m (6562-ft.) ASL
Storage temperature and
humidity
-10°C to 50°C (14 to 122°F), 80%RH or less (non-condensating)
Operating temperature and
humidity
0°C to 40°C (32 to 104°F), 80%RH or less (non-condensating)
Dielectric strength
1.69kV AC for 15s, Cutoff current 10 mA, between all power terminals and protective
ground, interfaces, and measurement jacks
Applicable Standards
Safety
EMC
EN61010
EN61326 Class A
EN61000-3-2
EN61000-3-3
Effect of radiated radio-frequency electromagnetic field: 3%f.s. at 10V/m
Effect of conducted radio-frequency electromagnetic field: 2%f.s. at 3 V
Summary of Contents for RM3542
Page 2: ......
Page 6: ...Contents iv ...
Page 16: ...Operating Precautions 10 ...
Page 26: ...1 3 Screen Organization 20 ...
Page 32: ...2 3 Turning the Power On and Off 26 ...
Page 46: ...3 8 Confirming Faulty Measurements 40 ...
Page 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Page 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Page 90: ...7 3 Printing 84 ...
Page 172: ...9 10 Device Compliance Statement 166 ...
Page 190: ...11 4 Disposing of the Instrument 184 ...
Page 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Page 214: ...Index Index 4 ...
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......