9.7 Message Reference
144
(10)Measurement Settings
Set and Query Delay 2 (after applying measurement current)
Syntax
Command
:RESistance:DELay2
<Range>
,
<Delay 2 time>
Query
Response
:RESistance:DELay2?
<Range>
<Delay 2 time>
Command
:LPResistance:DELay2
<Range>
,
<Delay 2 time>
Query
Response
:LPResistance:DELay2?
<Range>
<Delay 2 time>
<Range>
=
(:RESistance) RNG100MIL/ RNG1000MIL/ RNG10/ RNG100/ RNG1000/ RNG10K/
RNG100K/ RNG1000K/ RNG10MEG/ RNG100MEG
(:LPResistance) RNG1000MIL/ RNG10/ RNG100/ RNG1000
<Delay 2 time>
= 0 to 0.1 (seconds)
Example
:RES:DEL2 RNG1000MIL,0.003
:RES:DEL2? RNG10
0.005
Set and Query Integration Time (no. of power line cycles)
Syntax
Command
:RESistance:NPLCycles
<Range>
,
<Speed>
,
<Integration time
(NPLC)>
Query
Response
:RESistance:NPLCycles?
<Range>
,
<Speed>
<Integration time (NPLC)>
Command
:LPResistance:NPLCycles
<Range>
,
<Speed>
,
<Integration time
(NPLC)>
Query
Response
:LPResistance:NPLCycles?
<Range>
,
<Speed>
<Integration time (NPLC)>
<Range>
=
(:RESistance) RNG100MIL/ RNG1000MIL/ RNG10/ RNG100/ RNG1000/ RNG10K/
RNG100K/ RNG1000K/ RNG10MEG/ RNG100MEG
(:LPResistance) RNG1000MIL/ RNG10/ RNG100/ RNG1000
<Speed>
= SLOW/ MEDium/ FAST
<Integration time (NPLC)>
= 0.01 to 6
Description
Sets the integration time as an integer multiple of power line cycles (NPLC).
Note
•
For high-resistance measurements, hum noise is suppressed and stable measured values are
acquired by setting an integer multiple of power line cycles.
•
When converting to seconds, values over 100 ms are set to 100 ms.
•
When querying the integration time, the instrument’s internal variable is converted and output.
So in some cases, the returned value may not match the set value.
Summary of Contents for RM3542
Page 2: ......
Page 6: ...Contents iv ...
Page 16: ...Operating Precautions 10 ...
Page 26: ...1 3 Screen Organization 20 ...
Page 32: ...2 3 Turning the Power On and Off 26 ...
Page 46: ...3 8 Confirming Faulty Measurements 40 ...
Page 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Page 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Page 90: ...7 3 Printing 84 ...
Page 172: ...9 10 Device Compliance Statement 166 ...
Page 190: ...11 4 Disposing of the Instrument 184 ...
Page 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Page 214: ...Index Index 4 ...
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......