9.7 Message Reference
131
9
(5) Comparator
Query Comparator results
Syntax
Query
Response
:CALCulate:STATistics:LIMit?
<Hi count (NR1)>,<IN count (NR1)>,<Lo count (NR1)>,<Measurement fault count
(NR1)>,<Out-of-range count (NR1)>
Example
:CALC:STAT:LIM?
1516,9310,737,16,5
Query Standard Deviation
Syntax
Query
Response
:CALCulate:STATistics:DEViation?
<
n (NR3)>,<
n-1 (NR3)>
Example
:CALC:STAT:DEV?
0.0159E-3,0.0161E-3
Query the Process Capability Indices
Syntax
Query
Response
:CALCulate:STATistics:CP?
<Cp (NR2)>,<CpK (NR2)>
Example
:CALC:STAT:CP?
0.86,0.14
Set and Query Comparator Settings
When making comparator settings by commands, the measurement range is not automatically selected.
Comparator State
Syntax
Command
:CALCulate:LIMit:STATe
<1/ 0/ ON/ OFF>
Query
Response
:CALCulate:LIMit:STATe?
<ON/ OFF>
Example
:CALC:LIM:STAT ON
Beeper State
Syntax
Command
:CALCulate:LIMit:BEEPer
<OFF/ HI/ LO/ HL/ IN>
Query
Response
:CALCulate:LIMit:BEEPer?
<OFF/ HI/ LO/ HL/ IN>
Example
:CALC:LIM:BEEP HL
Summary of Contents for RM3542
Page 2: ......
Page 6: ...Contents iv ...
Page 16: ...Operating Precautions 10 ...
Page 26: ...1 3 Screen Organization 20 ...
Page 32: ...2 3 Turning the Power On and Off 26 ...
Page 46: ...3 8 Confirming Faulty Measurements 40 ...
Page 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Page 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Page 90: ...7 3 Printing 84 ...
Page 172: ...9 10 Device Compliance Statement 166 ...
Page 190: ...11 4 Disposing of the Instrument 184 ...
Page 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Page 214: ...Index Index 4 ...
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......