9.7 Message Reference
128
(2) Measurement-Related
Set and Query the Resistance Measurement Method
Syntax
Command
[:SENSe:]FUNCtion
<RESistance/ LPResistance>
Query
Response
[:SENSe:]FUNCtion?
<RESISTANCE/ LPRESISTANCE>
<RESISTANCE>
= Resistance measurement
<LPRESISTANCE>
= Low-Power Resistance measurement
Example
FUNC LPR
Selects the Low-Power Resistance measurement.
FUNC?
RESISTANCE
The Resistance measurement has been selected.
Note
[:SENSe:] may be omitted.
Set and Query the Range Setting
Low-Power Resistance Measurement Range
Syntax
Command
[:SENSe:]LPResistance:RANGe
<Expected measurement value>
Query
Response
[:SENSe:]LPResistance:RANGe?
<Measurement Range
(NR3)>
<Expected measurement value>
= 0 to 1200
<Measurement Range (NR3)>
= 1000.00E-3/ 10.0000E+0/ 100.000E+0/ 1000.00E+0
Description
Command
Enter the expected measurement value. The instrument is set to the most suitable
range for measuring the given numerical value data.
Query
Queries the measurement range setting.
Example
LPR:RANG?
10.0000E+0
Low-Power Resistance measurement has been set to the 10
range.
Resistance Measurement Range
Syntax
Command
[:SENSe:]RESistance:RANGe
<Expected measurement value>
Query
Response
[:SENSe:]RESistance:RANGe?
<Measurement Range (NR3)>
<Expected measurement value>
= 0 to 120E+6
<Measurement Range (NR3)>
= 100.000E-3/ 1000.00E-3/ 10.0000E+0/ 100.000E+0/
1000.00E+0/ 10.0000E+3/ 100.000E+3/ 1000.00E+3/ 10.0000E+6/ 100.0000E+6
Description
Command
Enter the expected measurement value. The instrument is set to the most suitable
range for measuring the given numerical value data.
Query
Queries the measurement range setting.
Example
RES:RANG 95
Sets the Resistance measurement to the 100
range.
Summary of Contents for RM3542
Page 2: ......
Page 6: ...Contents iv ...
Page 16: ...Operating Precautions 10 ...
Page 26: ...1 3 Screen Organization 20 ...
Page 32: ...2 3 Turning the Power On and Off 26 ...
Page 46: ...3 8 Confirming Faulty Measurements 40 ...
Page 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Page 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Page 90: ...7 3 Printing 84 ...
Page 172: ...9 10 Device Compliance Statement 166 ...
Page 190: ...11 4 Disposing of the Instrument 184 ...
Page 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Page 214: ...Index Index 4 ...
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......