4.13 Maintaining Measurement Precision (Self-Calibration Function)
65
4
To maintain measurement precision, the instrument self-calibrates every ten minutes to compen-
sate for internal circuit offset voltage and gain drift. This function cannot be disabled.
During self-calibration, the subsequent measurement is delayed for about 6 PLC + 10 ms (PLC =
Power Line Cycles) for internal circuit compensation.
Self-Calibration Timing
Supplied power 50 Hz: 130 ms, 60 Hz: 110 ms
• When the timing of self-calibration overlaps with a measurement, self-calibration is postponed until the end
of measurement.
• When a trigger signal (measurement start signal) is applied during self-calibration, the start of the triggered
measurement is postponed until self-calibration is finished.
• Self-calibration executes automatically after changing comparator or measurement speed settings.
• During self-calibration, measurement current and the Contact Improver current are inhibited.
This function automatically compensates for offset voltage resulting from thermal emf or internal
instrument bias. (OVC: Offset Voltage Compensation)
See:
"Appendix 2 Effect of Thermal emf" (p. A2)
The following value is known to be a true resistance value from
R
P
(>0), the value measured with
current flowing in the positive direction, and
R
N
(<0), the value measured with current flowing in the
negative direction.
Offset voltage compensation is automatically enabled in the following conditions, and cannot be modified or
disabled:
• When a range from 100 m
Ω
to 10
Ω
is selected.
• Low Power Resistance Measurement (LOW POWER ON)
When the test object is inductive, some delay [DELAY2] is required (p. 44) to allow adequate current flow
before starting measurement.
4.13 Maintaining Measurement Precision
(Self-Calibration Function)
4.14 Compensating for Thermal EMF Offset
(Offset Voltage Compensation - OVC)
R
P
R
N
–
2
-----------------------
Summary of Contents for RM3542-50
Page 1: ......
Page 2: ......
Page 6: ...Table of Contents iv...
Page 26: ...1 3 Screen Organization 20...
Page 32: ...2 3 Turning the Power On and Off 26...
Page 48: ...3 8 Confirming Faulty Measurements 42...
Page 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Page 84: ...5 7 Initializing Reset 78...
Page 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Page 206: ...11 4 Disposing of the Instrument 200...
Page 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Page 230: ...Index Index 4...
Page 231: ...HIOKI E E CORPORATION...
Page 232: ......
Page 233: ......
Page 234: ......