4.4 Setting the Measurement Integration Time Option
48
The integration time can be optionally set for each range by selecting FAST, MED, or SLOW mea-
surement speed.
Integration time can be set in ms or PLC
*
units.
* PLC = Power Line Cycle, where one PLC is the time equivalent to one period of the power line waveform.
At 50 Hz, one PLC = 1/50
th
of a second, and at 60 Hz, one PLC = 1/60
th
of a second.
PLC setting units are useful where measurements may be affected by power line noise (high- or low-resistance measure-
ments)
*1. Low-power resistance measurement (p. 28)
*2. Two measurements are made within the above integration time.
4.4 Setting the Measurement Integration Time
Option
Default setting
Range
LOW POWER
*1
: OFF
LOW POWER
*1
: ON
Integration time
OVC
Integration time
OVC
FAST
MED
SLOW
FAST
MED
SLOW
100 m
Ω
0.5 ms 5.0 ms 1 PLC ON
*2
−
−
−
−
1000 m
Ω
0.3 ms 2.5 ms 1 PLC ON
*2
0.5 ms 5.0 ms 1 PLC ON
*2
3
Ω
0.3 ms 2.5 ms 1 PLC ON
*2
0.5 ms 5.0 ms 1 PLC ON
*2
10
Ω
0.3 ms 2.5 ms 1 PLC ON
*2
0.5 ms 5.0 ms 1 PLC ON
*2
100
Ω
0.3 ms 3.0 ms 1 PLC OFF 0.3 ms 2.5 ms 1 PLC ON
*2
300
Ω
0.3 ms 3.0 ms 1 PLC OFF 0.3 ms 2.5 ms 1 PLC ON
*2
1000
Ω
0.3 ms 3.0 ms 1 PLC OFF 0.3 ms 2.5 ms 1 PLC ON
*2
10 k
Ω
0.3 ms 3.0 ms 1 PLC OFF
−
−
−
−
30 k
Ω
0.3 ms 3.0 ms 1 PLC OFF
−
−
−
−
100 k
Ω
0.5 ms 3.0 ms 1 PLC OFF
−
−
−
−
300 k
Ω
0.5 ms 3.0 ms 1 PLC OFF
−
−
−
−
1000 k
Ω
1.5 ms 5.0 ms 1 PLC OFF
-
-
-
-
3 M
Ω
1.5 ms 5.0 ms 1 PLC OFF
-
-
-
-
10 M
Ω
2.5 ms 1 PLC
1 PLC OFF
-
-
-
-
30 M
Ω
2.5 ms 1 PLC
1 PLC OFF
-
-
-
-
100 M
Ω
1 PLC
2 PLC
4 PLC OFF
-
-
-
-
1
Open the Basic Settings screen.
2
Open the Measurement Settings Screen.
The Basic Settings screen appears.
The Measurement Settings Screen
appears.
[MEAS SETTINGS]
(The settings for the current
measurement range are displayed.)
1
Selection
2
Summary of Contents for RM3542-50
Page 1: ......
Page 2: ......
Page 6: ...Table of Contents iv...
Page 26: ...1 3 Screen Organization 20...
Page 32: ...2 3 Turning the Power On and Off 26...
Page 48: ...3 8 Confirming Faulty Measurements 42...
Page 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Page 84: ...5 7 Initializing Reset 78...
Page 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Page 206: ...11 4 Disposing of the Instrument 200...
Page 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Page 230: ...Index Index 4...
Page 231: ...HIOKI E E CORPORATION...
Page 232: ......
Page 233: ......
Page 234: ......