9.7 Message Reference
135
9
Messages specific to the RS-232C or GP-IB interface are denoted by
or
.
(1) System Data Command
(2) Internal Operation Command
Shared Commands
Query the Device ID (Identification Code)
Syntax
Query
∗
IDN?
Response
<Manufacturer's name>, <Model name>, <0>, <Software version>
Example
HIOKI,RM3542A,0,V1.00
The Device ID is HIOKI RM3542A, 0, software version: 1.00.
Note
The response message has no header.
Initialize Device
Syntax
Command
∗
RST
Description
Command
Resets the instrument to its initial state.
Note
•
The communications protocol is not initialized.
•
Initialized settings are not backed up.
•
The following methods are available if backup is needed:
Method1. Use the
:SYSTem:BACKup
command
Method2. Use the
:SYSTem:RESet
command
Execute Self-Test and Query the Result
Syntax
Query
∗
TST?
Response
<0 to 7 (NR1)>
Description
Perform the instrument self-test and return the result as a numerical value from 0 to 7.
Returns zero when no error occurs.
Example
∗
TST?
4
A non-volatile memory error occurred.
Correct measurement may not be possible. Obtain repair before further use.
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
unused
unused
unused
unused
unused
non-volatile
memory
RAM
ROM
Summary of Contents for RM3542-50
Page 1: ......
Page 2: ......
Page 6: ...Table of Contents iv...
Page 26: ...1 3 Screen Organization 20...
Page 32: ...2 3 Turning the Power On and Off 26...
Page 48: ...3 8 Confirming Faulty Measurements 42...
Page 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Page 84: ...5 7 Initializing Reset 78...
Page 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Page 206: ...11 4 Disposing of the Instrument 200...
Page 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Page 230: ...Index Index 4...
Page 231: ...HIOKI E E CORPORATION...
Page 232: ......
Page 233: ......
Page 234: ......