Setting the Reduction Factor
4.6 Setting the Reduction Factor
You can set the magnitude of the correction applied during iterative calculation of interface
resistance (i.e., the reduction factor) to a value from 0.0001 to 1.0000.
1
2
3
4
1
Choose
[Settings]
.
2
Choose
[Analysis]
.
3
Enter the desired value in the
[Reduction factor]
field.
Valid setting range
Initial value
Format
0.0001 to 1.0000
1.0000
Decimal with 4 significant digits to the right of
the decimal point (0.0000)
4
Click
[Apply]
or
[OK]
to accept the settings. (Clicking
[OK]
will cause you to return to the
main screen.)
What is the reduction factor?
The reduction factor is a parameter for adjusting the magnitude of the correction applied
during iterative calculation of interface resistance. Use of a small value results in a more stable
convergence process but incurs longer calculation times. When convergence is stable, it is
recommended to use the initial setting of 1.0000. The setting can be changed as necessary
4
Customizing Measurement Conditions
Summary of Contents for RM2610
Page 2: ......
Page 12: ...8 Operating Precautions...
Page 22: ...18 USB License Key and Screens RM2612 Resistance Calculation Software...
Page 54: ...50 Checking Measurement Results...
Page 72: ...68 Setting Administrator Mode...
Page 110: ...106 Under what circumstances do contact errors occur...
Page 128: ...124 Dimensional Drawings 10 2 Dimensional Drawings RM2611 Electrode Resistance Meter Unit mm...
Page 130: ...126 Dimensional Drawings RM9005 Connection Cable Unit mm...
Page 134: ...130 Index...
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ......