41
Entering Electrode Sheet Information
3.5 Entering Electrode Sheet Information
In order to calculate electrode resistance, you will need to enter information about the electrode
sheet being measured. Analysis cannot be performed without this information.
IMPORTANT
If the operating mode is set to
[Analysis]
, the electrode sheet information stored in the specified potential
data for analysis file will be used instead of the electrode sheet information on the screen.
1
2
4
3
1
Enter the composite layer thickness in the
[Composite layer thick. [μm]]
field.
Valid setting range
Initial value
Format
0.0001 μm to 1000 μm
50
Decimal or exponential with 5 significant digits
2
Enter the collector layer thickness in the
[Collector layer thick. [μm]]
field.
Valid setting range
Initial value
Format
0.0001 μm to 1000 μm
20
Decimal or exponential with 5 significant digits
3
Enter the collector resistivity in the
[Collector resistivity [
Ω
cm]]
field.
Valid setting range
Initial value
Format
1.0000E-10 to 1.0000E+5
2.7000E-06*
Decimal or exponential with 5 significant digits
*Volume resistivity of aluminum
Volume resistivity (reference values)
You can choose between aluminum and copper from the pull-down menu.
Aluminum
2.7000E-06
Copper
1.7000E-06
3
Basic Measurement
Summary of Contents for RM2610
Page 2: ......
Page 12: ...8 Operating Precautions...
Page 22: ...18 USB License Key and Screens RM2612 Resistance Calculation Software...
Page 54: ...50 Checking Measurement Results...
Page 72: ...68 Setting Administrator Mode...
Page 110: ...106 Under what circumstances do contact errors occur...
Page 128: ...124 Dimensional Drawings 10 2 Dimensional Drawings RM2611 Electrode Resistance Meter Unit mm...
Page 130: ...126 Dimensional Drawings RM9005 Connection Cable Unit mm...
Page 134: ...130 Index...
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ......