How Should I Configure the Settings for a Collector with an Undercoat? Additionally, Is It Possible to Evaluate an
Und Collector Sheet That Lacks a Composite Layer?
8.7 How should I configure the settings for a
collector with an undercoat?
Additionally, is it
possible to evaluate an und collector
sheet that lacks a composite layer?
When the collector has an undercoat, it is recommended to measure the sheet by ignoring the
thickness of the undercoat when you enter measurement conditions.
Because the thickness of the undercoat is negligible compared to that of the composite layer and
collector layer, with the result that the effect of current flowing in the direction of the undercoat
surface can be ignored, the resistance value obtained by adding together the interface resistance
between the collector and the undercoat, the resistance of the undercoat itself, and the interface
resistance between the undercoat and the composite will be output as the interface resistance.
Since the undercoat layer is thin, it would be difficult for the instrument to evaluate the interface
resistance between the undercoat and the collector on a two-layer sheet made up of just an
undercoat and collector. Please evaluate the interface resistance based on the above interpretation
after applying the composite layer.
Interface
resistance
Undercoat layer
Interface resistance between
undercoat and composite
Interface resistance between
collector and undercoat
Composite layer
Collector layer
Figure: Diagram showing a cross-section of an electrode using a collector with an undercoat layer
8
FA
Q
Summary of Contents for RM2610
Page 2: ......
Page 12: ...8 Operating Precautions...
Page 22: ...18 USB License Key and Screens RM2612 Resistance Calculation Software...
Page 54: ...50 Checking Measurement Results...
Page 72: ...68 Setting Administrator Mode...
Page 110: ...106 Under what circumstances do contact errors occur...
Page 128: ...124 Dimensional Drawings 10 2 Dimensional Drawings RM2611 Electrode Resistance Meter Unit mm...
Page 130: ...126 Dimensional Drawings RM9005 Connection Cable Unit mm...
Page 134: ...130 Index...
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ......