66
Integrating Current and Power
Before starting integration
1
Adjust the clock.
See “6.1 Checking and Changing Settings” (p. 115).
2
Set the integration mode.
See “Integration mode” (p. 69).
3
Set the necessary control times.
See “Integration measurement while using the time control function” (p. 70).
Set the time settings to
[OFF]
when performing integration manually or with an external signal.
4
When saving data on a USB flash drive or generating D/A output, configure the
recording and D/A output settings.
See “7.1 USB Flash Drive” (p. 119) and “7.3 Saving Measured Data” (p. 123).
How to start/stop integration and resetting integrated values
These operations can be performed using the instrument’s control keys, external signals, or
communications.
Always reset integrated values when changing settings.
When the integration control is set to
[All Channel]
(Lit up in green)
1
Press the
START/STOP
key.
Integration starts.
The key lights up in green.
The integration state indicator will turn green.
(Lit up in red)
2
Press the
START/STOP
key.
Integration stops.
The key lights up in red.
The integration state indicator turns red.
(Off)
3
Press the
DATA RESET
key to reset integrated values.
The
START/STOP
key light is turned off.
The integration state indicator will be colorless.
When the timer control or real time control setting is used, integration will
stop automatically at the set end time.
HIOKI PW8001A961-00
Summary of Contents for PW8001-01
Page 2: ...HIOKI PW8001A961 00...
Page 6: ...Contents iv HIOKI PW8001A961 00...
Page 56: ...50 Checking Connections HIOKI PW8001A961 00...
Page 120: ...114 Power Calculation Method HIOKI PW8001A961 00...
Page 162: ...156 Controlling Integration with External Signals HIOKI PW8001A961 00...
Page 236: ...230 U7005 15MS s Input Unit HIOKI PW8001A961 00...
Page 258: ...252 Index HIOKI PW8001A961 00...
Page 259: ...HIOKI PW8001A961 00...
Page 260: ...HIOKI PW8001A961 00...
Page 261: ...HIOKI PW8001A961 00...
Page 262: ...www hioki com HIOKI PW8001A961 00...