109
Peak Hold Function
5.4 Peak Hold Function
Pressing the
PEAK HOLD
key places the instrument in the peak hold state. Only parameters
whose values exceed the past peak value are refreshed. This function is useful to thoroughly
capture phenomena characterized by instantaneously large values, for example rush current.
During peak hold operation, the
PEAK HOLD
key is lit up in red, and the
[PEAK HOLD]
icon on
the screen’s operating status indicator appears.
See “Common screen display” (p. 23).
Peak hold
Refresh
Refresh
Displayed value
Internally-retained
measured value
Peak hold starts.
(Lit up in red)
Peak hold is released.
Time
Display value
When the past peak value is exceeded, the displayed value for that parameter is refreshed.
Measurement continues internally.
Canceling the peak hold state
Press the
PEAK HOLD
key again during the peak hold operation to cancel the peak hold state.