![Hioki PW3390 Instruction Manual Download Page 25](http://html.mh-extra.com/html/hioki/pw3390/pw3390_instruction-manual_2137353025.webp)
PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
25/107
Set and Query Time
Syntax
Command
:CLOCk “Year Data”, “Month Data”, “Day Data”, “Hour Data”, “Minute
Data”, “Second Data”
Query
:CLOCk?
Response
“Year Data”, “Month Data”, “Day Data”, “Hour Data”, “Minute Data”, “Second
Data”
Year Data: 2000 - 2079 (can be set 00
– 79)
Month Data: 01- 12
Day Data: 01 - 31
Hour Data: 00 - 23
Minute Data: 00 - 59
Second Data: 0
Description
Command
Sets the time of the clock in the main instrument.
Query
Returns the time setting of the main instrument as NRI numerical values.
Example
Command
:CLOC 17,12,25,12,30,0
Set as 2017 December 25
th
12:30:0
Query
:CLOC?
Response
:CLOCK 2017,12,25,12,30,45 (when HEADER is ON)
2017,12,25,12,30,45 (when HEADER is OFF)
Note
The instrument can interpret days of the month as well as leap years, so specifying an
improbable date will lead to an error.
Always set 0 for the second data.
Summary of Contents for PW3390