![Hioki PW3390 Instruction Manual Download Page 11](http://html.mh-extra.com/html/hioki/pw3390/pw3390_instruction-manual_2137353011.webp)
PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
11/107
2. Command Reference (Standard Command)
Clear Standard Event Status Register (SESR) (except Output Queue)
Syntax
Command
*CLS
Example
Clear Event Register.
(SESR)
Note
・
No effect on Output Cue.
Read Standard Event Status Register (SESR)
Syntax
Query
*ESR?
Example
Return SESR Contents as NR1 numerical values 0-255.
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
PON
URQ
CME
EXE
DDE
QYE
RQC
OPC
PON: Power-On Flag. Set to 1 when the power is turned on, or upon recovery from an outage.
URQ:User Request. Unused.
CME:Command error. (The command to the message terminator is ignored.)
This bit is set to 1 when a received command contains a syntactic or semantic error:
• Program header error
• Incorrect number of data parameters
• Invalid parameter format
EXE: Execution Error
This bit is set to 1 when a received command cannot be executed for some reason.
• The specified data value is outside of the set range
• The specified setting data cannot be set (Invalid data format)
• Cannot be executed when another function is in operation (during hold, integration,
etc.)
DDE:Device-Dependent Error
This bit is set to 1 when a command cannot be executed due to
some reason other
than a command error, a query error or an execution error.
• Execution is impossible due to an internal instrument fault
QYE:Query Error (the output queue is cleared)
This bit is set to "1" when an abnormality occurs in processing an output queue.
• When the data overflows the output queue
RQC: Controller privilege request. Unused
OPC: Operation Complete. Unused
Example
Response *ESR 32 (when HEADER is ON)
32 (when HEADER is OFF)
Summary of Contents for PW3390