2.5 Supplying the Power
34
Power can be supplied from measurement lines using the PW9003 Voltage Line
Power Adapter (option).
Supplying Power from Measurement Lines
(Using the PW9003 Voltage Line Power Adapter)
To avoid electric shock or a short-circuit, observe the follow-
ing precautions:
• When using the PW9003 Voltage Line Power Adapter, never
wire the instrument to measurement lines with a voltage
exceeding 240 V.
• Make all connections after turning the PW9003 Voltage Line
Power Adapter’s power switch off. Making connections with
the power switch in the on position is extremely dangerous
and may cause sparks between measurement lines and the
metal tips on the voltage cords.
• Never wire the instrument to lines that include a high-fre-
quency component other than a commercial frequency, for
example the secondary side of an inverter.
• Before connecting the instrument to measurement lines,
verify that the voltage and frequency of the measurement
lines being used are 100 to 240 VAC and 50/60 Hz, respec-
tively. Use of the instrument outside the specified supply
voltage range may result in equipment damage or an electri-
cal accident.
• The instrument’s maximum rated power is 13 VA. When
using the PW9003 Voltage Line Power Adapter, do not do so
on the secondary side of a voltage transformer (VT or PT).
Avoid using an uninterruptible power supply (UPS) or DC/AC
inverter with rectangular wave or pseudo-sine-wave output to
power the instrument. Doing so may damage the instrument.
• Use of the PW9003 Voltage Line Power Adapter allows the
Z1006 AC Adapter to be connected to a CAT III (300 V) circuit
without sacrificing safety.
• The fuse is housed in the PW9003 Voltage Line Power Adapter.
If the power does not turn on, the fuse may be blown. If this
occurs, a replacement or repair cannot be performed by custom-
ers. Please contact your authorized Hioki distributor or reseller.
Summary of Contents for PW3360-20
Page 2: ......
Page 66: ...3 8 Verifying Correct Wiring Wiring Check 58 ...
Page 104: ...7 2 Settings That Can Be Added to Quick Set Settings 96 ...
Page 120: ...8 8 Formatting the SD Memory Card or Internal Memory 112 ...
Page 166: ...11 4 Outputting a Pulse Signal 158 ...
Page 210: ...13 4 Disposing of the Instrument 202 ...
Page 218: ...A8 ...
Page 222: ...Index Index 4 ...
Page 223: ......
Page 224: ......