3.11 When PEAK OVER, o.r, or the Unit Indicator Flashes
114
These lamps will light up if the voltage input or current input waveform peak value exceeds the figures listed
below. At this time, the displayed data is not accurate.
• Voltage input waveform peak value: ±600% of the voltage range
When using the 300 V, 600 V, or 1,000 V range, ±1,500 Vpeak
• Current input waveform peak value: ±600% of the current range
When using the 20 A or 50 A range, ±100 Apeak
3.11 When PEAK OVER, o.r, or the Unit Indicator
Flashes
3.11.1 If the PEAK OVER U or PEAK OVER I Lamp
Lights Up
Error display
Status
Solution
PEAK OVER U
When greater than or equal
to ±1,500 Vpeak
Stop measurement immediately, deactivate power to measurement
lines, and disconnect wires.
When less than ±1,500
Vpeak
The internal circuit is not operating properly. Switch to a range for
which the
PEAK OVER U
lamp does not light up.
See:
"3.2.4 Selecting the Voltage and Current Ranges" (p.48)
PEAK OVER I
When greater than or equal
to ±100 Apeak
Stop measurement immediately, deactivate power to measurement
lines, and disconnect wires.
When less than ±100 Apeak
The internal circuit is not operating properly. Switch to a range for
which the
PEAK OVER I
lamp does not light up.
See:
"3.2.4 Selecting the Voltage and Current Ranges" (p.48)
e.g.)
Summary of Contents for PW3336-01
Page 2: ......
Page 6: ...Contents iv...
Page 16: ...Operating Precautions 10...
Page 32: ...1 4 Measurement Workflow 26...
Page 44: ...2 7 Turning Off the Instrument 38...
Page 181: ...5 6 Wiring Specifications 175 5 3 phase 4 wire 3P4W Current sensor...
Page 182: ...5 6 Wiring Specifications 176...
Page 199: ...Appendix 4 Dimensional Diagram A13 e g PW3337 03 Appendix 4 Dimensional Diagram Unit mm...
Page 200: ...Appendix 4 Dimensional Diagram A14...
Page 203: ......
Page 204: ......
Page 205: ......
Page 206: ......