3.10 Other Functions
109
The instrument continuously measures instantaneous, maximum, and minimum values. The display can be
switched to these values with the
key.
• You can measure waveform peak values, for example device rush current (peak value hold).
• You can measure minimum values for measurement parameters (minimum value hold).
• You can measure maximum values for measurement parameters (maximum value hold).
Displaying maximum values
Displaying minimum values
Returning to the instantaneous value display
Press
while displaying minimum values (while the
MIN
lamp is lit up) to return to the instantaneous
value display.
Clearing maximum and minimum values
Pressing
and
(
DATA RESET
) will clear the maximum and minimum values and restart their
measurement. Maximum and minimum values are also cleared, and their measurement restarted, at the start
of integration.
3.10.2Displaying Peak, Minimum, and Maximum
Values (Maximum Value Hold)
1
Press
to place the instrument
in the shift state and then press
.
2
The
MAX
lamp will turn on, and the
display will switch to the maximum
value display.
1
Press
while the instrument is
displaying maximum values.
2
The
MIN
lamp will turn on, the dis-
play will switch to the minimum
value display.
Summary of Contents for PW3336-01
Page 2: ......
Page 6: ...Contents iv...
Page 16: ...Operating Precautions 10...
Page 32: ...1 4 Measurement Workflow 26...
Page 44: ...2 7 Turning Off the Instrument 38...
Page 181: ...5 6 Wiring Specifications 175 5 3 phase 4 wire 3P4W Current sensor...
Page 182: ...5 6 Wiring Specifications 176...
Page 199: ...Appendix 4 Dimensional Diagram A13 e g PW3337 03 Appendix 4 Dimensional Diagram Unit mm...
Page 200: ...Appendix 4 Dimensional Diagram A14...
Page 203: ......
Page 204: ......
Page 205: ......
Page 206: ......